APSGF Sterilizing-Grade Liquid Filter
Glass fiber is used as the pre-filtration membrane and PES is used as the final filtration membrane. It is an asymmetric polyethersulfone filter cartridges suitable for high-viscosity fluids.
Recommended Applications:
Request a Quote
Description
Features
- Asymmetric
- High capacity
- High flow rate
- Low Differential Pressure
Applications
- High viscosity fluid pre-filtration
Quality
- 100% Integrity testing in manufacturing
- Each filter is fully traceable with unique serial number
- Manufactured in a facility which adheres to ISO 9001:2015 Practices
Flow Rate Characteristics

Specifications
Materials of Construction
Membrane | Hydrophilic polyethersulfone (PES), Glass fiber (GF) |
Support | Polypropylene (PP) |
Core/Cage/End Caps | Polypropylene (PP) |
End Cap Inserts | Polybutylene terephthalate (PBT) |
Seal Material | Silicone / EPDM / Viton FEP/ PFA encapsulated O-rings |
Operating Conditions
Max. Operating Temperature |
80°C |
Max.Operating Pressure |
0.69 MPa@25°C 0.40 MPa@60°C 0.24 MPa@80°C |
Max. Differential Pressure (Forward) |
0.69 MPa@25°C 0.40 MPa@60°C 0.24 MPa@80°C |
Max. Differential Pressure (Reverse) |
0.30 MPa@25°C 0.10 MPa@80°C |
Filtration Area
Outer Diameter | Membrane Pore Size | Area/10" |
69 mm | 0.22 μm | 0.58 m² |
69 mm | 0.68 μm | 0.34 m² |
69 mm | 0.8 μm | 0.46 m² |
69 mm | 1.0 μm | 0.38 m² |
Integrity Test Standards @10inch,20°C
Membrane Pore Size | Bubble Point |
0.22 μm | ≥ 0.27 MPa (water), Air test |
Sterilization
In-line steam sterilization | Up to 25 cycles (135 °C for 30 min and differential pressure < 30 kPa per cycle) |
Autoclave | Up to 25 cycles (130 °C for 30 min per cycle) |
Regulatory Compliance
- Autoclaved filter effluent meets the USP<788> requirement of particulate matter in large volume injection.
- Component materials meet the criteria for a "Non-fiber-releasing filter" as defined in 21 CFR 210.3(b)(6).
- Aqueous extraction from a cartridge contains less than 0.25EU/ml as determined by Limulus Amebocyte Lysate (LAL), meeting requirements of USP<85>.
- Meet the requirement of USP <87> In Vitro Cytotoxicity Test.
- Component materials meet the requirements of the current USP<88> for plastic class VI-121°C.
- All component materials meet the FDA Indirect Food Additive requirements cited in 21 CFR 177-182.
- Based on the current information from our suppliers, all component materials used in the manufacture of this product are animal-free.
End Cap

Ordering Information
Outer Diameter | Type | Membrane Pore Size | End Cap | Nominal Length | Seal Material | Industry |
69mm | APSGF | 0022 0.22 µm | DOE Double Open End | 05 5 inch | S Silicone | P |
0065 0.65 µm | HTF 222/Fin | 10 10 inch | E EPDM | |||
0080 0.8 µm | HSF 226/Fin | 20 20 inch | V Viton | |||
0100 1.0 µm | HTCG 222/Flat | 30 30 inch | P FEP/PFA encapsulated O-rings | |||
HSCG 226/Flat | 40 40 inch |
* Example:APSGF0022DOE05SP (APSEA,0.22µm,Double open end,5 inch,Silicone,Pharmaceutical)