Suitable for particle retention and sterilizing filtration requirements in various gas environments.Excellent resistance to gamma irradiation, enabling used as a vent filter for single-use bags or culture flasks. Meets HEPA H13 filtration standard, achieving 99.97% removal efficiency for particles ≥0.3um. High dirt-holding capacity extends the service life of the filter.
Recommended Applications:
Request a Quote

Description

Features

  • Compact size for confined spaces 
  • High gas/air flow rate with reduced differential pressure
  • The hydrophobic glass fiber acts as a liquid barier, preventing aqueous solution penetration

Specifications

Filter Membrane Glass Fiber(GF)
Capsule Housing Polymethyl Methacrylate (PMMA)
Vent/Drain 1/4" HB
Pore Size 1.0 μm
Effective Filtration Area (EFA) 18.5 cm²
Bacterial Aerosol Retention Filter efficiency of Brevundimonas diminuta (ATCC®19146)>99.999%
Phage Aerosol Retention Filter efficiency of E. bacteriophage PhiX174(ATCC®13706-B1)>99.998%
0.3 μm Particle Retention ≥99.97 @ 2.83L/min
Gas Permeability ≥30 L/min @ 7 kPa
Compatible with Gamma Irradiation Sterilization Can withstand gamma irradiation with the dose of not less than 50 kGy.
The recommended irradiation dose is 25-45 kGy.
Sterile Disc Filter Gamma irradiated in sterile package and can not be re-sterilized.
Bacterial Endotoxin Aqueous extraction of autoclaved filter contains < 0.25 EU/mLas determined by Amebocyte Lysate, USP <85>.
USP <88> Biological Reactivity All the plastic component materials meet the FDA lndirect Food Additive reauirements cited in21 CFR 177-182.

Ordering Information